Power Requirements - GE AKTA avant Operating Instructions Manual

Preparative chromatography system
Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

4 Installation
4.1 Site preparation

4.1.4 Power requirements

4.1.4
Power requirements
Introduction
This section describes the electrical power requirements for the ÄKTA avant instrument.
Electrical power requirements
The following table specifies the power requirements.
Parameter
Supply voltage
Frequency
Transient overvoltages
Max power consump-
tion
Number of sockets
Type of sockets
Location of sockets
Quality of power
The mains power supply must be stable and conform to specifications at all times to
ensure reliable operation of the ÄKTA avant instrument. There should be no transient or
slow changes in average voltage outside the limits specified above.
60
Requirement
100 to 240 V AC ±10%
50/60 Hz
Overvoltage category II
800 VA
1 socket per instrument, up to 3 sockets for computer
equipment
EU or US plugs. Grounded mains sockets, fused or protect-
ed by equivalent circuit breaker.
Maximum 2 m from the instrument (due to length of mains
cable). Extension cables can be used if required.
WARNING
Protective ground. The product must always be connected to a
grounded power outlet.
ÄKTA avant Operating Instructions 29101556 AB