Diagnostic Mode

In Diagnostic Mode, the AreaRAE Plus/Pro provides raw counts for sensor, battery, and other
readings, as well as a list of installed sensors and information about them (expiration date, serial
number, etc.). Most of these screens are useful only to service technicians. Many allow access for
changing settings.
Diagnostic Mode can only be accessed at startup time. In Diagnostic Mode, readings are
displayed in raw counts instead of parts per million (ppm) or other units of measure.

Enter Diagnostic Mode

1.
With the instrument turned off, press and hold [Y/+] and [MODE] until the AreaRAE
Plus/Pro starts up.
2.
After startup, enter the password when prompted (default is "0000") and press [MODE].
Once Diagnostic Mode is active, step from screen to screen by pressing [N/-].
Adjusting Alarm LEDs & Buzzer
The buzzer, green LEDs, red LEDs, and backlight can be turned on or off in Diagnostic mode. In
addition, you can adjust the repetition rate and loudness of the buzzer and the repetition rate
and brightness of the LEDs.
1.
Start with the instrument turned off. Hold down the [Y/+] and [MODE] keys until the
AreaRAE Plus/Pro starts.
2.
When you see the password screen, input your 4-digit password, and then press [MODE].
3.
Once you have entered Diagnostic Mode, press [N/-] until the "Alarm" screen is displayed.
4.
Press [MODE] to step through the menu items; press [Y/+] to make a change, and press
[N/-] to advance to the next menu.
Important! When you are done, exit Diagnostic Mode and test the instrument before actual use.

Adjusting LCD Contrast

The display's contrast can be adjusted in Diagnostic mode.
1.
Start with the instrument turned off. Hold down the [Y/+] and [MODE] keys until the
AreaRAE Plus/Pro starts.
2.
When you see the password screen, input your 4-digit password, and then press [MODE].
3.
Once you have entered Diagnostic Mode, press [N/-] until the "LCD Contrast" screen is
displayed.
4.
Press [Y/+] to increase the value; press [MODE] to decrease the value, and press [N/-] to
advance to the next menu.