Replacing The Sensor Compartment Cover - Honeywell AreaRAE Plus User Manual

Wireless multi-gas, multi threat detector
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

6.
Replace the Lamp Housing Cap.
Note: Always fully calibrate the AreaRAE Plus/Pro after replacing the sensor module and/or
lamp.

Replacing the Sensor Compartment Cover

Before replacing the sensor compartment cover, inspect the large O-ring gasket for damage. If it
is damaged, discard it and use a new one.
1.
Place the O-ring gasket in the compartment. (Inspect the O-ring gasket for wear or
damage first. If necessary, replace it.)
2.
Press all the sensors firmly in place to ensure they are seated correctly.
3.
Set the sensor compartment cover on top of the compartment.
4.
Tighten the four screws. Note: Do not overtighten.
AreaRAE
112
User Manual