Honeywell AreaRAE Plus User Manual page 12

Wireless multi-gas, multi threat detector
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

AVERTISSEMENT: Comme condition de la certification, un connexion filaire ne peut pas être
faite via le port de communication lorsque le détecteur de gaz est dans la zone dangereuse à
moins que la zone soit connue pour ne pas être dangereuse.
WARNING: Li-Ion rechargeable battery: Only use approved battery pack: W01-3007-000 or W01-
3007-100.
AVERTISSEMENT: Batterie rechargeable Li-Ion: Utilisez uniquement des batteries approuvé:
W01-3007-000 ou W01-3007-100.
Equipment which is intended for use in explosive atmospheres and which has been assessed
and certified according to international regulations may be used only under specified
conditions. The components may not be modified in any way.
i.
Connection must only be made to the Vdc input jack and to the USB communication port
when the instrument is in non-hazardous location or unless the area is known to be non-
hazardous.
ii.
The battery pack shall only be charged or replaced in a non-hazardous location.
iii.
The sensors shall only be replaced in a non-hazardous location.
iv.
The external connection to the PGM 65xxx shall be in accordance to control drawing W01
0901-CTL. The external alarms that are to be connected to PGM 65xxx must be suitable
for the hazardous location where they are mounted, and supplied from Class 2 power
supply or equivalent. The wiring method shall be to local electrical code and installation
are subject to acceptance by authority having jurisdiction.
The appropriate regulations for service and repair must be properly observed during such
activities.
PGM-65XXX contains a connector for mounting of a wind sensor – RAEMet. This connector has
a set of entity parameters that match the input entity parameters of RAEMet:
Uo: 5.2V; Po: 0.788W; Lo: 198µH; Co: 1000µF.
Keep all ports covered when they are not in use. This keeps moisture and debris out of the ports
and contributes to the instrument's safety.
AreaRAE
USE IN HAZARDOUS AREAS
11
User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Arearae proPgm-6520Pgm-6560

Table of Contents