Sensor Locations
Location Sensor type(s)
1
4R+ EC --- O2 (this is the preferred slot)
2
4R+ EC or 4R+ PID or 4R+ CO2 (For best results, do not use HF or HCl sensors in this slot.)
3
4R+ EC (avoid placing the O2 sensor in this slot)
4
4R+ EC
5
4R+ EC or LEL (For best results, do not use HF or HCl sensors in this slot.)
6
4R+ EC (avoid placing the O2 sensor in this slot)
7
7R+ sensors
8
7R+ sensors
Note: The AreaRAE Plus/Pro does not allow duplicate sensors (that is, two or more of the same
kind, including, for example, a 4R+ PID and a 7R+ PID) to operate simultaneously. Therefore, if
duplicate sensors are inserted in the slots, when the instrument is turned on, it automatically
senses the redundancy and only allows operation of one of the sensors, selecting the sensor in
the location with the lowest number. The exception is the PID sensor: If both a 7R+ PID sensor
and a 4R+ PID sensor are detected, then the 7R+ PID sensor is selected for operation.
If you are removing a sensor and not replacing it, its slot cannot be left empty. A "dummy"
sensor must be installed in the empty slot.
AreaRAE
WARNING!
103
User Manual