General Alarms
Message
Condition
HIGH
Gas exceeds "High Alarm" limit
OVR
Gas exceeds sensor's measurement range
Gas exceeds electronic circuit's maximum range. If alarm persists in
MAX
clean air, it indicates a fault sensor.
LOW
Gas exceeds "Low Alarm" limit*
TWA
Gas exceeds "TWA" limit
STEL
Gas exceeds "STEL" limit
Crossed
pump icon
Inlet blocked or pump failure
flashes
"Lamp"
PID lamp failure
flashes
Empty
battery icon
Low battery
flashes
CAL
Calibration failed, or needs calibration
True sensor reading is below zero, even though a zero reading is
NEG
shown for the sensor. If alarm persists in clean air, it indicates a fault
sensor.
* For oxygen, "low alarm limit" means a concentration is lower than the low alarm limit.
AreaRAE
134
Alarm
Indications
3
beeps/flashes
per second
3
beeps/flashes
per second
3
beeps/flashes
per second
2
beeps/flashes
per second
1 Beep/flash
per second
1 Beep/flash
per second
3
beeps/flashes
per second
3
beeps/flashes
per second
1 flash, 1
beep per
minute
1 beep/flash
per second
1 beep/flash
per second
User Manual