Removing/Cleaning/Replacing Sensor Modules

WARNING! Do not replace sensors in hazardous locations.
All sensors are located inside the sensor compartment in the upper half of the AreaRAE
Plus/Pro. They are accessed by removing the cover that is held on by four screws.
1.
Turn off the instrument.
2.
Remove the four screws holding the sensor compartment.
3.
Remove the cover. The sensors are plugged into the slots.
4.
Gently lift out the desired sensor module with your fingers.
5.
Install the replacement sensor. It can only go into its slot one way. The connector inside
the AreaRAE Plus/Pro and the indexing guides are good visual indicators of how to set
the sensor into position. Make sure the indexing keys are aligned and press the sensor
into place to ensure it is seated firmly. Note that the sensors are also in different sizes,
and one size cannot be substituted for another.