Honeywell AreaRAE Plus User Manual page 59

Wireless multi-gas, multi threat detector
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

Note: Dotted line indicates automatic progression.
Multi Sensor Bump
Depending on the configuration of your AreaRAE Plus/Pro and the span gas you have, you can
perform a bump test simultaneously on multiple sensors.
The selected sensors are shown on the screen. With calibration gas connected to the instrument,
start a multiple bump test by applying gas to the instrument and pressing [Y/+]. If you do not
want to perform a multiple bump test, press [MODE].
Note: You can abort a multiple bump test by pressing [N/-] once testing has started.
When the Multi Bump test is done, a screen is shown, with the sensor names and either "Pass" or
"Fail" shown next to them.
Single Sensor Bump
This menu allows a bump test to be performed on an individual sensor of your choice.
Note: If a bump test icon (bottle with bottom portion not filled in) is shown next to any of the
sensors, it means that the sensor is due for a bump test.