Honeywell AreaRAE Plus User Manual page 60

Wireless multi-gas, multi threat detector
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

To perform a bump test on an individual sensor, follow these steps:
1.
At the Calibration Menu, select "Single Sensor Bump."
2.
Scroll down the list using [N/-], and then press [Y/+] to select a sensor to bump test.
3.
Install the "T" tube and connect it to a source of calibration gas.
4.
Verify that the displayed calibration value meets the concentration specified on the gas
cylinder.
5.
Start the flow of calibration gas.
Note: Dotted line indicates automatic progression.
6.
Press [Y/+] to start calibrating or wait for the bump test to start automatically.
7.
A countdown screen appears. You can abort the bump test at any time during the
countdown by pressing [N/-].
8.
If the bump test is not aborted, the display shows the sensor names and tells you whether
the bump test passed or failed, followed by the sensor readings.