Honeywell AreaRAE Plus User Manual page 141

Wireless multi-gas, multi threat detector
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

LEL Range, Resolution & Response Time
Range
0-100%
LEL Correction Factors
Compound
Acetone
Ammonia
Benzene
Butane, n-
Carbon monoxide
Cyclohexane
Ethanol
Ethyl acetate
Gasoline
Heptane, n-
Hexane, n-
Hydrogen
Isobutene
(Isobutylene)
Isopropanol
Methane
Methanol
Methyl ethyl ketone
Octane, n-
Propane
Propene
Toluene
Turpentine
Note: Values in bold type are confirmed with RAE Systems by Honeywell instruments. Others,
shown in italic, are calculated from diffusion models.
AreaRAE
Resolution
1 %
100% LEL (Vol %)
2.5
15
1.2
1.9
12.5
1.3
3.3
2
1.3
1.1
1.1
4
1.8
2
5
6
4.5
1
2.1
2
1.1
0.8
Response Time
15 sec
140
LEL CF*
1.9
1.0
2.1
1.9
1.3
2.1
1.8
2.4
2.6
2.5
2.1
1.0
1.6
2.2
1.0
1.6
2.2
2.7
1.4
1.6
2.4
3.0
User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Arearae proPgm-6520Pgm-6560