Record an ECG or Rhythm

Record a Full Disclosure ECG

Before you start this procedure, make sure:
• You purchase and enable the Full Disclosure option in the Options Manager.
• You enable the Full Disclosure option in the Setting screen.
1.
Start a new patient test.
2.
To view a Full Disclosure ECG, click the Full Disclosure tab on the Acquisition
screen.
• One lead of the Full Disclosure ECG displays. The Full Disclosure shows the
• The Full Disclosure ECG records for a maximum of 5 minutes. The recording
• The Full Disclosure screen displays 10 lines of ECG data and each line is of 30
• A notification message displays on the Acquisition screen after the Full
• The Full Disclosure screen displays the previous 5 minutes of Full Disclosure
3.
To change the lead, click Select Lead.
All the configured leads display. Select the single lead you want to display on the
screen and on the printed FD reports. If you connect the leads to a patient after
being fully disconnected for at least 30 seconds, all the data will be cleared from
the display.
The ECG recording restarts and the selected lead is applied to the Full Disclosure
waveform. All the previous recorded data will be cleared.
4.
To restart the Full Disclosure ECG, click Restart. All of the current waveform data
is deleted.
The message unsaved data will be lost. Confirm to proceed displays.
5.
To record a 10 seconds ECG when in the Full Disclosure screen, do the steps that
follow:
90
waveform from the left to right side of the screen.
stops after the 5 minutes is complete.
seconds.
Disclosure ECG records for 5 minutes.
ECG data.