Configure Display Formats Of Ecg And Rhythm Leads - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Field
Suppress
Abnormal
Statements
Screening
Mode
Reason
Statements
3.
Click Save.

Configure Display Formats of ECG and Rhythm Leads

Table 35: Default Lead Formats for Each Lead Set
Name
Adult 12 Lead
All
Leads
5864335-001-1
Action
Enable or disable
this setting.
Enable or disable
this setting.
Enable or disable
this setting.
No. of
Lead
Leads
Leads
Format
12
4x3
CH1 to CH12: I, II, III, aVR, aVL, aVF, V1,
V2, V3, V4, V5, V6
  MAC
5 A4/MAC
Description
If this setting is enabled, no abnormal or borderline
interpretative statements generate or display on the
report when you do as follows:
View a report in the preview and review screens
View a stored report in the review screen
Print a report
Send a report to a configured destination.
If this setting is disabled, abnormal and borderline
interpretative statements display on the report.
Default value: Disabled
If this setting is enabled, the device runs the 12SL
algorithm in high-specificity mode, where you will
not see certain lower-acuity statements in the
interpretation.
If this setting is disabled, the device runs the 12SL
algorithm in normal analysis mode and you see the
lower-acuity statements.
Default value: Disabled
You can select this setting only if you enable the
Screening Mode.
If this setting is enabled, reason statements generate or
display on the report when you perform the following:
View a report in the preview and review screens
View a stored report in the review screen
Print a report
Send a report to a configured destination.
If this setting is disabled, no reason statements
generate or display on the report.
Default value: Disabled
Default
Yes
5 A5/MAC
5 Lite
Configure Settings
Auto-
Digital
rhythm
Rhythm
No
Yes
 137