GE MAC 5 A4 Operator's Manual page 188

Resting ecg analysis system
Table of Contents

Advertisement

Configure Settings
Table 56: MUSE Server Settings for Orders
Field
Server URL
User Name
188
Action
Description
Enter a valid http
NOTE:
or https URL of
the MUSE server.
The URL must correspond to the MUSE system.
Allowed values: A valid http or https URL with defined
FQDN or IP address and a port number.
Use a Fully Qualified Domain Name (FQDN) or IP
address.
Add the port numeber after the URL with a colon (:)
specifier.
NOTE:
Determine which MUSE version you will connect to and
configure the URL:
Enter the MUSE
This field cannot be blank.
account user
Default value: No default value
name.
NOTE:
Allowed values:
Up to 128 characters
  MAC
5 A4/MAC
5 A5/MAC
You can secure some communication channels
with encryption and authentication. GE
Healthcare recommends that you use the
encrypted channels, not the unencrypted
channels. For example:
Upgrade the MUSE server 8.0/9.0 API3 version
to the MUSE NX API3 version, to have a secure
communication between the MAC 5 system
and the MUSE server.
Define the port number if it is not defined.
HTTP - 80
HTTPS - 443
MUSE system V8/V9: HTTP
NOTE:
MUSE system V8/V9 default port is 8100.
MUSE NX system: HTTPS
This is a MUSE account, not a Windows account.
a to z
A to Z
0 to 9
All special characters
5 Lite
 5864335-001-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mac 5 a5Mac 5 lite

Table of Contents