Ecg Specifications - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Item
Input current range
Input frequency
Table 121: Battery Specifications
Item
Type
Charge time
Battery capacity
Table 122: Other Input Device Specifications
Item
External USB barcode reader

ECG Specifications

Table 123: ECG Data Acquisition Specifications
Item
Signal input
Dynamic range
Common mode rejection
Input impedance
Patient leakage current
Detection of pacemaker pulse
5864335-001-1
Description
A @100V to 625mA @240V
780 mA @100V AC to 110 mA @240V AC
50/60 Hz ± 3 Hz
Description
Replaceable and rechargeable internal battery
Approximately 240 minutes from total discharge when device is off
(standby).
The system shall be able to acquire and print a single page 12-
lead Resting ECG report every fifteen minutes for a minimum
of three hours when the battery is installed and the battery is
fully charged (with five minutes auto standby enabled and all
accessories on except KISS), without AC connection.
Description
Fixed and variable length types
Symbologies: Code-128, PDF417, Code 39, Interleaved Code 2 of
5, and Data Matrix symbology for characters A-Z (upper case), a-z
(lower case), and 0-9 for all supported languages.
Description
Type CF, Defibrillation-Proof
Defibrillation protection: Per IEC 60601-2-25:2011
AC Differential ± 10mV, DC offset ± 600 mV
>125 dB (>100 dB with AC filter disabled)
>50 MΩ @ 10 Hz, defibrillator protected
<10 µA
Duration: 0.2 ms to 2.1 ms
Amplitude: 2 mV to 700 mV
Separation: 1 ms or greater
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
Technical Specifications
 335