Table 47: Configure a DCP Server Destination to Send Reports
Field Name
Destination
Name
Server Name
ADT Query
5.
Select Test Connection.
• If the test displays Test Successful, you have a successful connection to that
• If the test displays Test Failed, you do not have a connection to that
6.
Select Save.
7.
Repeat steps
8.
To edit an existing DCP server destination:
a)
5864335-001-1
Action
Enter the name of the DCP server destination
where the reports will be sent.
Enter the URL of the DCP server.
NOTE:
Make sure that you append '/SendTest'
to the URL. For example,
<IP_Address> or <Hostname>:<Port>/
SendTest.
Confirm that the URL for the server is
correct.
Confirm that the DCP server is running.
Make sure that you enable ADT for DCP
communication, configure the same IP
address for the DCP destination and the
MUSE Orders Server for remote query.
Enable or disable this setting.
destination. Select Add to save the destination.
destination. Troubleshoot the connection failure depending on the error, re-
test and add the connection.
to
to add more DCP server destinations.
2
6
Select anywhere in the row of the destination you want to modify to enable
the edit mode.
  MAC
http://
5 A4/MAC
5 A5/MAC
5 Lite
Configure Settings
Description
A user-defined value
up to 20 characters.
Allowed values:
A to Z
a to z
0 to 9
All special
characters
A user-defined value.
Allowed values:
A to Z
a to z
0 to 9
All special
characters
If this setting
is enabled, the
destination is
configured to perform
ADT query.
Default value: Disabled