GE MAC 5 A4 Operator's Manual page 139

Resting ecg analysis system
Table of Contents

Advertisement

3.
Select the Add icon
A new row is added to the lead format table.
4.
Configure the lead format as per the information in the table:
Table 36: Display Format Settings for ECG and Rhythm Leads
Field
Name
Number of
Leads
Lead Format
Leads
Default
5864335-001-1
to add a lead format.
Action
Description
Enter a name for
Allowed values:
your lead format
Up to 20 characters. Allowed values are:
setting.
A to Z
a to z
0 to 9
All special characters
Select the number
Default value:
of leads you want
For 12 leads: 12
to include in the
lead format.
Allowed values:
For 12 leads: 3, 6, 12
Select the layout
The different types of lead formats are as follows:
for the leads in
3 Leads: 1x3
columns by rows.
6 leads: 1x6, 2x3, or 2x3 Simult
12 leads: 2x6, 2x6 Simult, 4x3, 4x3 Simult
Simult refers to display all the leads at the same time.
You can add up to 10 new format entries.
Default value:
For 12 leads: 4x3
Allowed values:
For 12 leads: 4x3, 4x3 Simult, 2x6, 2x6 Simult
Select the leads
Default value:
in each channel
Adult 12 Lead:
that you want
to display in the
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
waveform for the
Allowed values:
selected lead set.
Adult 12 Lead:
I, II, III, aVR, -aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Enable or disable
If this setting is enabled, this lead set format is the
this setting.
default format for ECGs recorded on this device.
Default value: Disabled
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
Configure Settings
 139