Ecg Acquisition Errors - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Troubleshooting
Message
Battery not detected
Device date/time is incorrect.
Update.
Touchscreen failure
Cannot perform action while
acquiring ECG data

ECG Acquisition Errors

The table lists messages that you may encounter while acquiring ECG.
276
Cause
The battery is not detected by
the device.
The date and time set on the
device is not correct.
The touchscreen is not working. Contact your GE Healthcare
You have tried to perform
below actions while acquiring
an ECG or rhythm:
Start a test for a new
patient
Enter or edit patient
information (available for
rhythm)
Change speed, gain, or filter
(available for rhythm)
Access User Menu
Access Orders, Files, or
Queue list
Start ECG
Power off, lock, log out,
standby or privacy mode
  MAC
5 A4/MAC
5 A5/MAC
Solution
Perform below steps:
1.
Remove the battery.
2.
Re-insert the battery.
If the battery is still not
detected, replace the battery.
Contact your GE Healthcare
Service Support representative
if the error persists.
Select Adjust to set the correct
date and time. For more
information,
Date and Time on page
Service Support representative
if the error persists.
Perform only allowed actions.
5 Lite
seConfigure the
255e .
 5864335-001-1