Chapter 4. Programming
4.6.2c
Sound Speed Validation
When the speed of sound in your fluid does not equal published values, refer to
Figure 83 below and set the SOUND SPEED LEVEL by completing the following
steps:
1.
Click the SOUND SPEED LEVEL input control.
2.
Enter your actual sound speed in the units specified by your system UNITS
setting.
3.
Click the OK button to accept the new value.
Important:
You have now completed the programming of the PIPE , FLUID ,
TRANSDUCER
button to change to the measurement display or click the LOG DATA
button to begin logging data.
76
Figure 83: Sound Speed Validation
and PLACEMENT menus. Click on the GO TO MEASURE
®