Chapter 4. Programming
4.5.1 Programming the Transducer Parameters (cont.)
2.
Click on the TRANSDUCERS button and select your transducer model from
the drop down list. The drop-down list will vary, based on whether you
selected Wetted or Clamp-on as your transducer type. After you make
your selection, a screen similar to Figure 76 below will open.
3.
In the screen above, the PT900 has automatically entered the parameters
for your selected transducer. This data is accurate for all Standard
transducers, but you must edit this data with the correct information if you
selected Other as your transducer model.
a.
Open the FREQUENCY drop-down list and select the frequency at
which your transducer is designed to operate.
b. The TW parameter is the time the transducer signal spends traveling
through the transducer and the transducer cable. Enter the value
provided by GE with your transducer.
68
Figure 76: The Transducer Parameters
®
TransPort