Floor/Tabletop Mounting - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Installation

Floor/tabletop mounting

3-8
a.
If a connection to the ATS Unity Network IX network ethernet port is used,
configure a service laptop for direct ethernet connection with the ATS.
Service laptop direct connection on page C-2.
b.
If a serial connection to the ATS is used, connect the serial connection.
Serial connection on page 6-4.
2.
Start-up the ATS.
See Power up on page 3-14.
3.
Configure the ATS Unity Network IX network IP address settings.
ApexPro Channelized Telemetry System network name and IP address on page
5-4.
4.
Shut down the ATS.
See Safe shutdown/restart of the ATS on page 6-21.
with ATS installation.
1.
Remove the 5 center screws from the right side of the ATS.
2.
Remove the screws and/or the handle at the front right side of the ATS.
3.
Place the ATS on its left side on a sturdy surface. Position the stand so the holes
in the stand align with the screw holes in the right side of the ATS. Secure the
stand with the 5 screws removed in step 1.
4.
Finish securing the stand with the 2 front right side screws removed with the
handle in step 3.
ApexPro™ Telemetry
See
See
See Set the
Proceed
159A