Network Diagnostics - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

View hardware environmental parameter data via BIOS settings

Network diagnostics

Adaptors - Display adaptor and serial port information
2001989-203C
The Temperatures diagnostics page displays real-time ATS internal temperature
information. Temperatures must be between the High- and Low Limit levels
displayed.
NOTE
If an error message displays,
via BIOS settings on page
In Webmin, if you are unable to view the hardware status on fan speed, power, or
temperature, you can alternatively view the information via the BIOS settings.
1.
Log off Webmin.
2.
Follow the instructions for
3.
Restart the system. While the system is starting up, press the Del key until the
Advanced BIOS Features screen displays.
4.
Navigate to PC Health Status.
5.
Note the fan speed, power, and temperature information on the PC Health Status
screen.
1.
Log onto the Webmin service interface on page
2.
Select the Diagnostics tab.
ApexPro™ Telemetry
View hardware environmental parameter data
6-11.
Safe shutdown/restart of the ATS on page
4-7.
Troubleshooting
6-21.
601A
6-11

Advertisement