Parts lists, drawings and replacement

Hard drive

Replace the hard drive
7-18
Perform the following procedures to replace the hard drive and verify that the rebuild
procedure is complete:
Replace the hard drive on page 7-18
GE recommends that you replace the defective hard drive with a known blank,
formatted hard drive supplied as a FRU.
The top and bottom hard drives have different procedures. Refer to the appropriate
procedure:
Bottom hard drive on page 7-18
Top hard drive on page 7-19
Bottom hard drive
1.
Record the serial number of the new hard drive you are installing.
2.
7-3.
3.
Remove the chassis cover.
4.
Remove the drive bay.
5.
Using a phillips screwdriver, remove the 4 anchor screws securing the bottom
hard drive on both sides inside the drive bay.
6.
Slide the bottom hard drive out of the back of the drive bay.
ApexPro™ Telemetry
See Chassis cover on page 7-6.
See Drive bay on page 7-16.
018A
074A
2001989-203C