Raid Card - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Parts lists, drawings and replacement

RAID card

7-8
The FRU kit can contain either of the following RAID cards to replace the RAID card
in the ATS:
Part number
2021518-001
Promise FastTrak 150 RAID card initially released with ATS.
Compatible with all ATS OS Image versions.
2021518-002
Promise FastTrak 2300 RAID card released due to obsolescence of
or
original card.
2021518-003
Compatible with ATS OS Image v1.3 and greater.
1.
Look at the part number of the RAID card you are about to install in the ATS.
Promise FastTrak 150 RAID card (pn 2021518-001).
Promise FastTrak 2300 RAID card (pn 2021518-002).
Promise FastTrak 2300 RAID card (pn 2021518-003).
2.
Verify that the proper ATS OS image version is running.
image version on page 4-11.
3.
Using the table below, along with the ATS OS image version and RAID card part
number (identified in step 1), determine the appropriate action.
Kit containing
ATS OS image
part number
2021518-001
All
2021518-002/
1.3 or greater
2021518-003
2021518-002/
Earlier than 1.3
2021518-003
4.
Use the ATS Software Reload Instructions included in the ATS RAID FRU Kit to
reload the ATS to ATS OS image v1.3 or greater.
ApexPro™ Telemetry
Description
version
Replace the part. See step 5.
Replace the part. See step 5.
Software must be executed before replacing part.
See step 4.
See View the ATS OS
Action
2001989-203C

Advertisement

Table of Contents
loading

Table of Contents