Troubleshooting

Hardware monitoring utilities

Fans
Power
Temperature
6-10
Specific information associated with rectifying conditions indicated by an audible
alarm is located in the service status indicators section.
on page 6-8.
NOTE
If Webmin displays an error message when you select the Fans, Power, or
Temperature screens, this indicates that there is a communication error. The
environmental parameter data can be collected via BIOS settings.
hardware environmental parameter data via BIOS settings on page 6-11.
1.
Log onto the Webmin service interface on page
2.
Select the Diagnostics tab.
3.
Select Hardware > Fans.
The Fans diagnostics page displays the current fan speed. The current fan speed
must be faster than the Low Limit threshold.
NOTE
If an error message displays,
via BIOS settings on page
1.
Log onto the Webmin service interface on page
2.
Select the Diagnostics tab.
3.
Select Hardware > Power.
The Power diagnostics page displays real-time power supply status, and real-
time CPU and System power levels. Power levels must be between the High- and
Low Limit levels displayed.
NOTE
If an error message displays,
via BIOS settings on page
1.
Log onto the Webmin service interface on page
2.
Select the Diagnostics tab.
3.
Select Hardware > Temperatures.
ApexPro™ Telemetry
See Service status indicators
4-7.
View hardware environmental parameter data
6-11.
4-7.
View hardware environmental parameter data
6-11.
4-7.
009A
See View
2001989-203C