Honeywell VC4 User Manual page 160

4-point continuous monitor
Hide thumbs Also See for VC4:
Table of Contents

Advertisement

          Point to be configured - 1 byte
                    bits 0-1: Point #
                              00 = Point 1
                              01 = Point 2
                              10 = Point 3
                              11 = Point 4
          Gas Table # - 1 byte (0 = first table)
          Alarm Level 1 - 2 bytes
          Alarm Level 2 - 2 bytes
          20 mA Full Scale - 2 bytes
          Point ID - 20 bytes
Response packet from instrument:
          Command Code (0x59) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Status - 1 byte
                    0x00 = Point configured and verified
                    0x01 = Gas error
                    0x02 = Alarm 1 Err (L1 < LAL or L1 > FS)
                    0x04 = Alarm 2 Error (L2 < L1 or L2 > FS)
                    0x08 = 20 mA Error (< LAL or > FS)
                    0xff = Save problem
Set TWA Time - 0x5A
This command configures the TWA time output. You need to enter only the initial TWA output
time. The other two time factors are calculated automatically.
Command packet to instrument:
          Command Code (0x5A) - 1 byte
          TWA Time - 2 bytes
          Command Code (0x5A) - 1 byte
VC4™ Date -2 bytes
VC4™ Time - 2 bytes
VC4™
160
User Manual