Response Packet from instrument:
          Command Code (0x37) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes (If point is Disabled/Invalid/locked out, fill with zeroes)
          MDA Gas Abbr. - 6 bytes (not null terminated)
          Format Code - 1 byte
          Flow Rate - 2 bytes (current flow)
          TWA Start Date - 2 bytes
          TWA Start Time - 2 bytes
          TWA End Date -2 bytes
          TWA End Time - 2 bytes
          TWA Conc. - 2 bytes
          Last Conc. - 2 bytes (last reported conc.)
          Alarm Status - 1 byte (0 none, 1 or 2 level) (Fill for all conditions) Status - 1 byte
                    0x00 = Data valid
                    0x01 = Point Disabled (no data filled)
                    0x02 = Point Locked Out (no data filled)
                    0x04 = No TWA calculated
                    0x08 = No concentration available
                    0x10 = Alarm Simulation mode active
                    0xff = Invalid data
Get TWA Time - 0x38
This command queries the three TWA output time in a 24-hour format. Each of the TWA times
are eight hours apart.
Command packet to instrument:
          Command Code (0x38) - 1 byte
Response packet from instrument:
          Command Code (0x38) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          TWA time 1 - 2 bytes
          TWA time 2 - 2 bytes
          TWA time 3 - 2 bytes
VC4™