Response packet from instrument:
          Command Code (0x56) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time -2 bytes
          Return Status - 1 byte
                    0x00 = Counter is enable/disable
                    0x01 = No windows left
                    0x02 = Maintenance status exists (Low Chemcassette™)
                    0xff = Error in programming counter
Set Printer Configuration - 0x57
This command configures the printer for output.
Command packet to instrument:
          Command Code (0x57) - 1 byte
          Setup Status - 1 byte
                    bit 0: Printer port enable/disable
                    bits 1-2: Printer Report format
                              00 = Continuous (prints all conc. for each pt)
                              01 = Summary (prints alarms, faults, and TWA only)
                              10 = Compressed (prints conc. > LDL)
                              11 = Invalid
                    bits 3-5: Printer baud rate
                              000 = 1200
                              001 = 2400
                              010 = 4800
                              011 = 9600
                              100 = 19200
                    bit 6: printer hardware handshaking (flow control)
                              0 = Disabled
                              1 = Enabled
                    bit 7: Undefined
Response packet from instrument:
          Command Code (0x57) - 1 byte
VC4™ Date - 2 bytes
VC4™