Honeywell VC4 User Manual page 112

4-point continuous monitor
Hide thumbs Also See for VC4:
Table of Contents

Advertisement

Event
Description
ID
128
Ethernet initialization failed
129
File system of Analyzer is corrupted
130
*Reserved
131
Unable To Log event data
Software version mismatched
132
among Analyzers
This Chemcassette™ is nearing its
133
expiration date
134
Chemcassette™ Type Not Matched
135
Analyzer configuration failed
136
Point configuration failed
137
Could not start monitoring
138
Analyzer RTC not set correctly
139
*Reserved
140
*Reserved
141
*Reserved
142
*Reserved
143
Pump current overload
144
Temperature out of range
Possible Cause
Failed load the driver, Electronic problem
File system corrupted
File system corrupted
New analyzer was installed into the VC4™ rack that contains a different software revision
than the other analyzers
Chemcassette™ not used too long
Chemcassette™ with wrong gas family installed
Analyzer configuration failed
Point configuration failed
Instrument faults not cleared
Invalid RFID tag detected
All points disabled
Invalid Analyzer/Point Configuration
LIT in progress
Tape Advance Failure
Low voltage of the coin battery
RTC failure on the analyzer board
Line voltage issue
Fan failure
Line voltage issue
Resolution
Reboot the VC4™
Contact Honeywell Service
Contact Honeywell Service
Contact Honeywell Service
Upgrade analyzer Software
Replace Chemcassette™
Reinstall Configuration Profile
Reinstall Configuration Profile
Check alarm settings and reinstall Configuration Profile
Resolve the reported instrument faults and reset all faults
Reinstall Configuration Profile or Replace Chemcassette™
Reinstall Configuration Profile or enable runtime disabled
points
Reinstall Configuration Profile
Wait until LIT test is complete and restart monitoring mode
"Instrument Faults" on the next page for Fault 233
Replace the coin battery
Replace Analyzer
Verify main line voltage
Check fans in pump module
Verify main line voltage