6.3 Maintenance Faults

A maintenance fault indicates the VC4™ system requires attention but is continuing to monitor.
Event
Description
ID
101
Flow is 70 cc/min less than nominal
102
Remaining Chemcassette™ is low
103
Analyzer out of monitor too long
104
Run time point disable time out
Sample pressures greater than
105
ambient
Flow is 70 cc/min more than
106
nominal
107
Chemcassette™ expired
108
No polls from HMI for 10 seconds
109
*Reserved
Possible Cause
Excessive point vacuum due to clog or kinked sample line
Sample line too long or ID too small
Poor gate seal
Supply vacuum insufficient (less than 7"Hg)
Condensation
Chemcassette™ thickness variation
Less than preset reminder on the Chemcassette™
Analyzer out of monitor
Out of Monitor time limit too short
Alarms were manually bypassed
Run Time Point Disable time limit too short
Point pressure above atmospheric pressure while in idle
Pressure sensor Calibration error, Defective sensor
Chemcassette™ thickness variation
Defective needle valve
Chemcassette™ expiration date reached
Chemcassette™ installed past its expiration date
Communications to HMI PC interrupted longer than 10 seconds
Resolution
Check sample line and line filter
Clean the sample line and replace filter
Correct sample line issue
Ensure sample line requirements are with specifications
Check nut on the optic block
Contact Honeywell Service
Contact Honeywell Service
Check internal lines for moisture
Advance the Chemcassette™
Adjust the needle valve to achieve 200cc/min
Contact Honeywell Service
Replace the Chemcassette™
Reset all alarms and faults, and then return analyzer to
Monitor Mode
Change time limit in Configuration Profile
Restore point to alarm active mode
Change time limit in Configuration Profile
Review sample line location
Confirm atmospheric conditions
Contact Honeywell Service
Advance the Chemcassette™
Adjust the needle valve to achieve 200cc/min
Contact Honeywell Service
Replace needle valve
Replace Chemcassette™
Replace Chemcassette™
Check Ethernet cable at rear of analyzer
Check Ethernet hub connection and operation
Check Ethernet connection to HMI PC
Contact Honeywell Service