Event
Description
ID
238
Analyzer SW corrupted
239
Internal SW fault at analyzer
240
Pump failure
241
IO panel not responding to Analyzer
242
Internal fault at IO panel module
243
FAN fail
Possible Cause
Software installation failure
Internal SW operation failed
Line voltage issue
Pump failure
Circuit breaker tripped
Leakage in system
Cable issue
IO panel SW operation failed
IO panel SW operation failed
Fan blade gets stuck
Fan short circuit
Resolution
Re-install analyzer software
Replace Analyzer, Contact Honeywell service
Reboot the Analyzer. Contact Honeywell service
Verify main line voltage
Replace pump, Contact Honeywell service
Contact Honeywell service
Check tubing and fitting in system, Contact Honeywell service
Check IO panel cabling
Reboot the IO panel module. Contact Honeywell service
Reboot the IO panel module. Contact Honeywell service
Check fan blade, Contact Honeywell Analytics Service
Replace fan, Contact Honeywell Analytics Service