Honeywell VC4 User Manual page 159

4-point continuous monitor
Hide thumbs Also See for VC4:
Table of Contents

Advertisement

VC4™ Time - 2 bytes
          Status Flag - 1 byte
                    0x00 = Printer programmed
                    0x01 = Invalid report format
                    0xff = Printer programming error
Set Point Enable/Disable – 0x58
This command enables or disables points on the VC4™ monitor.
Command packet to instrument:
          Command Code (0x58) - 1 byte
          Point enable mask - 1 byte
                    bits 0-3: Point selection
                              0x01 = Point 1 enabeled
                              0x02 = Point 2 enabeled
                              0x04 = Point 3 enabeled
                              0x08 = point 4 enabeled
                    bits 4-7: Undefined
Response packet from instrument:
          Command Code (0x58) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Status - 1 byte
                    0x00 = Point enable updated and verified
                    0xff = Error, not saved
Set Point Configuration - 0x59
This command configures an individual point. The point can be configured even if it is disabled.
Command packet to instrument:
          Command Code (0x59) - 1 byte
VC4™
159
User Manual

Advertisement

Table of Contents
loading