Honeywell VC4 User Manual page 142

4-point continuous monitor
Hide thumbs Also See for VC4:
Table of Contents

Advertisement

Maintenance Status
                              0x01 = low flow point 1
                              0x02 = low flow point 2
                              0x04 = low flow point 3
                              0x08 = low flow point 4
                              0x10 = low Chemcassette™ tape
                              0x20 = Maint. Relay
                              0x40 = Instr. Fault Relay
Get Idle Time - 0x32
This command inquires about how long the unit can be left out of analysis before setting an
Instrument Fault. Idle time is used to notify operators that the unit is not monitoring. An idle
time of 0 disables this option.
Command packet to instrument:
          Command Code (0x32) - 1 byte
Response packet from instrument:
          Command Code (0x32) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Idle Time - 1 byte (0 disabled, 1-45 min)
          Status - 1 byte
                    0x00 = No errors in reading
                    0xff = Error in reading
Get Date & Time - 0x33
This command retrieves the current date and time from the unit.
Command packet to instrument:
          Command Code (0x33) - 1 byte
Response packet from instrument:
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Status - 1 byte
                    0x00 = Date and Time READ
                    0xff = Read problem
VC4™
142
User Manual