Battery care and maintenance

ADTS Touch battery pack
The battery unit contains a Lithium Ion battery which requires no user maintenance.
Prolonged exposure to temperature extremes may significantly reduce battery lifetime. For maximum
lifetime, avoid prolonged periods where the battery is exposed to temperatures outside the range
-30°C to +45°C.
The recommended storage temperature range is 5°C to 21°C (41°F to 98.8°F).
The battery is removable. The manufacturer makes the following safety recommendations. The User
should not:
Short circuit the battery.
Immerse the battery in any liquid.
Disassemble or deform the battery.
Expose to, or dispose of the battery in a fire.
Subject the battery to excessive physical shock or vibration outside the specified limits for the
ADTS.
Use a battery that appears to have suffered abuse.
The battery unit is sealed for life, so leakage of electrolyte is not expected. If leakage of electrolyte from
the battery is observed, stop using the battery immediately and avoid contact with the electrolyte. If
skin or clothes come into contact with the electrolyte, wash immediately with soap and water. If elec-
trolyte comes into contact with the eyes, wash the eyes thoroughly with water and consult a Doctor
immediately.