Suggested two-point calibration adjustment points
Calibration points are nominal pressures to allow for slight differences due to primary standard
weights and actual corrected pressure calculations.
- Ps: two point cal points = 1st point is 1128mbar (FS) followed by 92mbar (suggest to be done in this
order).
- Pt: two point cal points = 1st point 92mbar followed by 1997mbar (FS) (suggest to be done in this
order).
Suggested calibration check points
Ps (Calibration check points)
56.39 mbar
71.71 mbar
92.00 mbar
178.74 mbar
314.85 mbar
465.63 mbar
696.82 mbar
843.07 mbar
1013.25 mbar
1128.03 mbar
-
-
Note: ~means approximately
a. *ADTS552/3/4, ** ADTS552/3/4 EALT option.
Table 4-4: Calibration check points
(~65,000 ft)**
(~60,000 ft)*
(~55,000ft)
(~41,000ft)
(~29,000ft)
(~20,000ft)
(~10,000ft)
(~5,000ft)
(~0ft)
(~-3,000ft)
-
-
a
C
ALIBRATION PROCESS
Pt (Calibration check
points)
56.39 mbar**
71.71 mbar*
92.00mbar
178.74mbar
314.85mbar
465.63mbar
696.82mbar
843.07mbar
1013.25mbar
1128.03mbar
1500mbar
1997mbar