GE ADTS 542 Instruction Manual page 85

Air data test set
Table of Contents

Advertisement

T
OOLS
Save/Recall ADTS settings
ADTS Manuals
Customer documents
Calibration (Calibrate sensors)
To access this function you will be required to enter your PIN. This function is used to set new
corrected values for the sensors based on the outcome of the calibration procedures detailed in
section Calibration Description. The Calibration sub-menu contains the following items:
- Sensor
Opens the Calibration check sub-menu showing the current values for Ps and Pt:
To enter new correction values for Ps:
1)
Tap the Ps panel. The Ps sensor correction panel opens.
2)
Follow the on-screen instructions. Tap the "tick" icon. A numbered keypad opens.
3)
Perform the same procedure to enter new correction values for Pt.
- Change CAL PIN
When selected, this function allows you to enter a new PIN code. To change your PIN:
1)
Tap the "Change CAL PIN" panel. A numbered keypad opens.
2)
Enter the new PIN number and tap the "tick" icon. You will be asked to confirm the change.
3)
Enter the new PIN again and tap the "tick" icon again. The numbered keypad closes and the new
PIN is now active.
4)
Tapping the "Cross" icon cancels the action and closes the numbered keypad without changing
the PIN.
Calibration (Software upgrade)
To access this function you will be required to enter your PIN. This function is used to implement soft-
ware upgrades for the ADTS Tap and the ADTS following installation of updated software. The "Cali-
bration" sub-menu contains the following items:
83
Save ADTS settings
Recall ADTS settings
Delete ADTS settings
Copy all files from USB
Copy all files to USB
Restore last ADTS settings
Set ADTS touch defaults
Set ADTS defaults