Access Summaries - Honeywell VPR100 User Manual

Table of Contents

Advertisement

Online Operation Using Menus

7.4 Access Summaries

Summaries are available for all points, alarms and diagnostics in the instrument.
Display alarm summary
Alarms are set up as part of the instrument configuration procedure (Section 4.13). Up to 16 alarms can be
configured.
An alarm can be assigned to any analog data point (Analog Input, Analog Output, or Calculated Value) and
can be one of these types.
Alarm type
HIGH
LOW
DEV (Deviation)
LDEV
HDEV
HRATE (High Rate)
LRATE (Low Rate)
An alarm programmed with delay will not activate before its delay time. An alarm programmed with
hysteresis will clear after its hysteresis delay.
The alarm will remain active as long as the conditions causing it remain. When the conditions no longer
exist, the alarm will be "cleared" automatically. "Clear" means that the indicators for the particular alarm
on all displays will be removed and the alarm will be removed from the Alarm Summary list.
240
Function
Alarm when input value > setpoint value.
Alarm when input value < setpoint value.
Alarm when input value deviates above or below compare point
value by an amount > setpoint value.
Low Deviation. Alarm when input value deviates below compare
point value by an amount > setpoint value.
High Deviation. Alarm when input value deviates aboveow compare
point value by an amount > setpoint value.
Alarm when input value increases at rate > setpoint value, in units
per minute. Negative rate setpoints are processed as positive
values. May take up to 30 seconds to activate.
Alarm when input value decreases at rate > setpoint value, in units
per minute. Negative rate setpoints are processed as positive
values. May take up to 30 seconds to activate.
VPR & VRX – User Manual
12/00

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vrx100Vrx150

Table of Contents