Keywords – logfile.txt
"Starting Log System"
-
EAErrorHandler
-
"Inside Next"
-
Power Button Press
Info
........
Info
Info
Print Button Press
Info
Scan state, Probe Connected, etc.
-
System Hardware ID
-
Front End Card Crack Configuration
acqHWCap(200); HW Rack Configuration Valid: Ichiro Phase 2 - CW (2A/2C), DRX MLA4
(3E), GTX (2E), GRLY (2C), MLP PS (1C)
OR
Info
2B), DRX MLA4 (3B), GTX (2C), GRLY (2C), MLP PS (1A)
7 - 270
PROPRIETARY TO GE
. 7
EV
First line written whenever application starts
Indicates serious error has occurred. Usually seen several times for each failure.
Never seen except when Diags are run.
Error
; EAErrorHandler(3660); Error from HouseKeeping: GFI Processing
Exception: SubType 166:
inside next [0093ebc0,00941ff8> vector! (b=56, s=280).
65576, no of beams received: 65548. Severity : NonFatalSW
; UserEventLog(304); ButtonPress: name=PowerBut, value=0
; SysExitDlg(1912); System Exit Dialog Displayed
; SysExitDlg(1912); System Exit Dialog - Shutdown Selected
; UserEventLog(168); ButtonPress: name=PrintBut1, value=0
Look for ESStatus
Info
; EScan.View(1288); ESStatus(0,Live): SetState RunState("Live" ->
"Freeze") Modes("2DHAR") Probe("cla_15C-D") Appl("Abdomen")
Stored on the backplane – Should not be ZEROs!
Info
; EchoRoot(1288); HW Serial Number = 0x005E3250 = 6173264, 96348US8
; acqHWCap(1904); HW Rack Configuration Valid: Ichiro Phase 1 - NoCW (2A/
Section 7-9 - Troubleshooting Tips
LOGIQ E9 P
scanseq Halting scan: s_current_iq_addr(0093ebc0)
S
M
ROPRIETARY
ERVICE
ANUAL
No of firings done: