System boots, with hardware error - Lambda PS only

Table 7-54
System boots, with hardware
error
7 - 210
PROPRIETARY TO GE
. 7
EV
System boots, with hardware error - Lambda PS only
Symptom
The latest Firmware Version is
1.21. Check Firmware Version
on the PS label or check the
SysCfg.Csv file in the Alt D
Section 7-9 - Troubleshooting Tips
Task / Conclusion
Replace the PS
logs.
Check if 24V is present
LOGIQ E9 P
ROPRIETARY
The firmware in the PS may have been
corrupted and the system is not able to control
the power supply. The logfile.txt will report
USB errors since it cannot see the PS USB
device and will prevent scanning.
Example:
......Monday,Jul 25 14:57:19,2011; Error ;
acqTransLayer(288); CMicrochipUsbIf::Open
Cannot open output pipe (last error
0x00000000)
......Monday,Jul 25 14:57:19,2011; Info ;
ifmgr(208); 14:57:19
HandleEvent:
IMLoader(0), EAErrorHandler(2467),
(BSTR)HWCAP Invalid Hardware
Configuration, (BSTR)Invalid Hardware
Configuration(68) 6029.3ms }
NOTE: A long term solution is already in place
to prevent possible scenarios where the
firmware may get corrupted. This solution has
been implemented in production.
Manufacturing believes that only 50 units
shipped with this old firmware may be
susceptible to this problem.