Log file changes in R4

Date format changes: Adding more time resolution down to milliseconds
Sequence was improved to reflect real time occurrence of events and errors
• Before: event will display the time when entering the log
• Now: log entry shows real time of occurrence.
Date/ me
2012/08/29 13:00:09.769
2012/08/29 13:00:09.771
2012/08/29 13:00:09.771
Fields added for trending and future troubleshooting tools to read logs
A
B
C
R3
Friday,Apr 20 11:33:04,2012; info ; StartLoader(0x0); ************ START **************
Friday,Apr 20 11:33:04,2012; info ; StartLoader(0x0); StartLoader Version: "V 3.0.0.5"
Friday,Apr 20 11:33:05,2012; info ; StartLoader(0x0); Turning Rack Power On
Friday,Apr 20 11:33:07,2012; info ; StartLoader(0x0); Starting Echoloader
Friday,Apr 20 11:33:07,2012; info ; StartLoader(0x0); CreateProcess returned success: ProcessId = 0x774 (1908)
Friday,Apr 20 11:33:07,2012; info ; StartLoader(0x0); Echoloader process started
Friday,Apr 20 11:33:23,2012; info ; StartLoader(0x0); WMI Reports Processor "CPU0" ClockSpeed is 2133 (max 2133) NumberOfCores is 2 NumberOfLogicalProcessors is 2
Friday,Apr 20 11:33:23,2012; info ; StartLoader(0x0); MSR 408 reports highpart 0x082A082A lowpart 0x0600082A
Friday,Apr 20 11:33:23,2012; info ; StartLoader(0x0); PerformanceState 2090 (0x82A)
Friday,Apr 20 11:38:36,2012; info ; StartLoader(0x0); Power Button Detected
Friday,Apr 20 11:38:36,2012; info ; StartLoader(0x0); Power Button Ignore Delay Active (Exit Dialog Displayed)
Friday,Apr 20 11:38:53,2012; info ; StartLoader(0x0); Process stopped, exit code = 0/2 (Shutdown OS)
Friday,Apr 20 11:38:53,2012; info ; StartLoader(0x0); Shutdown OS
Friday,Apr 20 11:38:53,2012; info ; StartLoader(0x0); Terminating
Friday,Apr 20 11:38:53,2012; info ; StartLoader(0x0); Killing BlockWindowsKeys
R4
2012/08/21 08:55:10.881; info ;StartLoader.cpp(974); ************ START **************
2012/08/21 08:55:10.881; info ;StartLoader.cpp(976); StartLoader Version: "V 4.0.0.0"
2012/08/21 08:55:12.192; info ;StartLoader.cpp(162); Turning Rack Power On
2012/08/21 08:55:12.192; info ;StartLoader.cpp(168); Unable to open PS USB interface
2012/08/21 08:55:19.648; info ;StartLoader.cpp(1471); Starting Echoloader
2012/08/21 08:55:19.648; info ;StartLoader.cpp(1492); CreateProcess returned success: ProcessId = 0xE70 (3696)
2012/08/21 08:55:19.758; info ;StartLoader.cpp(1498); Echoloader process started
2012/08/21 08:55:20.834; info ;StartLoader.cpp(2405); WMI Reports Processor "CPU0" ClockSpeed is 3101 (max 3101) NumberOfCores is 4 NumberOfLogicalProcessors is 4
2012/08/21 08:57:25.197; info ;StartLoader.cpp(1680); Process stopped, exit code = 0/2 (Shutdown OS)
2012/08/21 08:57:25.197; info ;StartLoader.cpp(798); Shutdown OS
2012/08/21 08:57:25.213; info ;StartLoader.cpp(1894); Terminating
2012/08/21 09:00:46.866; info ;StartLoader.cpp(974); ************ START **************
7 - 260
PROPRIETARY TO GE
. 7
EV
Figure 7-167 Log file changes in R4
Because of this, sequential log entries may not be sequential in time.
Type Moduke
Message
ESStatus(0,Live): SetMode RunState("Live") Modes("<none>" -> "2D")
Info EScan.View(3844)
Probe("fla_11L-D") Appl("Stress")
Info
TPDebug:ActivateTab called for All value=false force=false
EScan.TabMgr(3844)
Info
TPDebug:These tabs are displayed:
EScan.TabMgr(3844)
Log message number
Log message ID (Internal Job identifier)
Where the message comes from in the Software code
Figure 7-168 Log file changes in R4 (cont.)
Internal job identifier
New date format
Section 7-9 - Troubleshooting Tips
LOGIQ E9 P
ROPRIETARY
A
B
C
6697
Views\ESView.cpp(7099)
1130
c5a3
ESTabMgr.cpp(2584)
1131
ac60
ESTabMgr.cpp(2585)
1132
BEP5 information
BEP6 information
S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents