D
5180483-100, R
IRECTION
7-9-18-5
Errors versus Logs (cont'd)
Bad Hardware
•
From logfile.txt:
Error
SubType 62:
temperature[37] RX2_TOP. Severity : FatalHW
•
System tried to read the temperature but got no or invalid response
•
Can be temperature or voltages
•
Voltage reading errors are usually NOT hardware failures, but single failures of the system to read
a voltage. Replace hardware only when repeated errors are seen.
•
Log\logfile.txt Example:
Error
; acqHWCAP(228); ********************************************
Error
; acqHWCAP(228); ***
Error
; acqHWCap(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ********************************************
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ***
Error
; acqHWCAP(228); ********************************************
Info
; DFEManager(228); HandleRunFlag: setting RunDfe and RunBlock flags to true!
Error ; EAErrorHandler(228); Error from acqHWCAP: Invalid Hardware Rack Configuration - Check logs. Severity :
FatalHW
Info
; DFEManager(228); HandleRunFlag: setting RunBlock flag to false!
Warning; EAErrorHandler(228); Stops scan
ONLY SHOWS WHAT WAS FOUND, NOT WHAT WAS BAD!
PROPRIETARY TO GE
. 7
EV
; EAErrorHandler(2168); Error from HouseKeeping: GFI Hardware Exception:
Monitoring : RFIMonitor::handleMonitoring() - Failed reading
HW RACK CONFIGURATION INVALID!!!
Board FuncRevs Provided
-----------------------
BP_FuncRev2A
(qty=1)
DRX_FuncRev3E
(qty=3)
GFI_FuncRev2A
(qty=1)
GRLY_FuncRev2C
(qty=1)
GRX_FuncRev2A
(qty=1)
GRX_FuncRev2B
(qty=1)
GTX_FuncRev2C
(qty=3)
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
***
***
***
***
***
***
***
S
M
ERVICE
ANUAL
7 - 265