GE LOGIQ E9 Service Manual page 418

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-4-6
Main Power Supply Logs (cont'd)
Examples of Main Power Supply Information at Startup - Normal Messages
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3056); TxPs Alarm source[System
(rack) supply]
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3056); TxPs TS On Strobe Event
discarded (root cause of HW watchdog stop will handle the event)
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3060); TxPs alarm status AC-
V[116.4 VAC] AC-Freq[60 Hz] ACErr[no] RackOn[48V ACinv]
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3060); +48V[48.02 Volts] +24V[3.64
Volts] +15V[0.00 Volts] -15V[-11.79 Volts] +12V[0.00 Volts] +6V[0.00 Volts] -6V[-1.86
Volts] +5V[4.96 Volts] --
TURNED OFF. (This is not considered a failure.)
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3060); inTemp[25 °C] outTemp[26
°C] +PMX V[27.11 Volts] -PMX V[-18.41 Volts] --
AFTER THEY WERE TURNED OFF. (This is not considered a failure.)
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3064); TxPs alarm info
source[Transmit supply 1] V+[0.32 Volts] V-[0.00 Volts] Power[0.00 Watts]
Current[0.000 Amps] Prgmd V[0.00 Volts]
Friday,Feb 12 07:27:52,2010; Info; acqHouseKeeping(3064); TxPs alarm info
source[Transmit supply 2] V+[0.32 Volts] V-[0.00 Volts] Power[0.00 Watts]
Current[0.000 Amps] Prgmd V[0.00 Volts]
Additional Examples of Main Power Supply Information at Startup - Normal Messages
Friday,Feb 12 07:26:39,2010; Info; acqHouseKeeping(3060); TxPs alarm status AC-
V[116.1 VAC] AC-Freq[60 Hz] ACErr[yes] RackOn[48V LV ACinv]
ACErr[yes] =
by battery operation message.)
[48V LV ACinv] =
Friday,Feb 12 07:26:39,2010; Info; acqHouseKeeping(3060); +48V[48.02 Volts]
+24V[24.07 Volts] +15V[15.00 Volts] -15V[-14.98 Volts] +12V[11.02 Volts] +6V[5.90
Volts] -6V[-6.03 Volts] +5V[4.95 Volts]
Friday,Feb 12 07:26:39,2010; Info; acqHouseKeeping(3060); inTemp[24 °C] outTemp[24
°C] +PMX V[0.00 Volts] -PMX V[0.00 Volts]
Friday,Feb 12 07:26:39,2010; Info; acqHouseKeeping(3064); TxPs alarm info
source[Transmit supply 2] V+[0.32 Volts] V-[0.00 Volts] Power[0.00 Watts]
Current[0.000 Amps] Prgmd V[0.00 Volts]
Friday,Feb 12 07:26:39,2010; Info; acqHouseKeeping(3056); TxPs Alarm source[System
(rack) supply]
Friday,Feb 12 07:26:39,2010; Info; acqTransLayer(3048); CUsbPsAdapter:Init Power
supply protocol version 4
PROPRIETARY TO GE
. 7
EV
type[Error] detail[TS On Strobe]
THESE VOLTAGES WERE MEASURED AFTER THEY WERE
(Power Supply was disconnected from AC. Not an indication of an error, unless followed
(Not an indication of an error, unless followed by battery operation message.)
type[No alarm] detail[TS On Strobe]
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
THESE VOLTAGES WERE MEASURED
S
M
ERVICE
ANUAL
7 - 215