GE LOGIQ E9 Service Manual page 184

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
Table 3-25
7.
To confirm a successful Storage Commitment, press F4 to inspect the spooler after you send images
to the Image Storage service. An entry of:
• Type: STC
• Contents: Value equals the number of images stored, plus one if a Structured Report was sent.
• Status: "Done" or "Success"
• Error: no indication confirms a successful storage commitment.
3 - 72
PROPRIETARY TO GE
. 7
EV
DICOM Storage Commitment Service Setup and Use
In the Dataflow tab, add your Storage Commitment service to the same dataflow that handles
your Image Storage service.
Storage Commitment includes "reverse negotiation" with the receiving end – the Image
Storage service. Basically, the receiver needs to know the AE Title, IP address and Port
Number of the LOGIQ E9. Refer to the documentation of the Image Storage service for setting
up Storage Commitment.
NOTE: The Store Commit and its Associated Storage (A) MUST BE in the same Dataflow!
Section 3-9 - Connectivity Setup and Tips
DICOM Job Spooler - STC identified
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents