Shear Wave Option Troubleshooting - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-15

Shear Wave Option Troubleshooting

Table 7-59
Shear Wave Control is not available on the
Touch Screen.
Performance is poor.
Fuse is blown or has blown while in Shear
Wave (the fuse IS NOT replaceable).
PROPRIETARY TO GE
. 7
EV
Shear Wave Option Troubleshooting
Symptom
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
Is Option Hardware installed properly and Option String present?
User will not be able to select the Shear Wave Control on the Touch Screen
(Control is not available).
Reboot to check if Shear Wave is still available.
While the LOGIQ E9 is OFF and Capacitor Pack is discharged, check Capacitor
Cable connections.
NOTE: Both Capacitor Pack Module LEDs (two modules, four LEDs total) should
be lit after few seconds of booting up (visible through the Capacitor Pack Cover).
BCM2 is only used during CW Doppler + Simultaneous.
If all Capacitors are NOT charging, the charge time will be affected. Even if the
LEDs are lit. At boot up the system should detect the failure and disable Shear
Wave.
S
ROPRIETARY
ERVICE
Action
M
ANUAL
7 - 239