Key Log Files (cont'd)
7-9-18-3-1
Logfile-TempMon.txt
Should be imported into Excel for easier reading
-
Key Items: DRX3_TOP
Friday,Mar 19 07:52:14,2010; Info
Friday,Mar 19 07:52:14,2010; Info
Temp
Out;DRX2_LM83;DRX2_BOT;DRX2_TOP;DRX2_DCDC;DRX2_GDIF;DRX2_1062;DRX3_LM83;DRX3_BOT;DR
X3_TOP;DRX3_DCDC;DRX3_GDIF;DRX3_1062;DRX4_LM83;DRX4_BOT;DRX4_TOP;DRX4_DCDC;DRX4_GDI
F;DRX4_1062;GFI_LM83;GFI_GFE;GFI_SDEM;GFI_BOT;GRX1_LM83;GRX1_TOP;GRX1_BOT;GRX2_LM83
;GRX2_TOP;GRX2_BOT;GTX2_TOP;GTX2_1062;GTX3_TOP;GTX3_1062;GTX4_TOP;GTX4_1062;EPS
Temp;Fan_Control;Rack Fan 1;Rack Fan 2;Rack Fan 3;Rack Fan
4;BEP_SYS_TEMPERATURE;BEP_CPU_TEMPERATURE;BEP_AUX_TEMPERATURE;BEP_SYS_FANSPEED;BEP_
CPU_FANSPEED;GFI External Temperature;GFI Internal Temperature
Friday,Mar 19 07:53:25,2010; Info
25;
27;
34;
33;
19.00;
Key
Values in file are preceded by:
NOT ALL VALUES IN LOG ARE PROPERLY MONITORED.
PROPRIETARY TO GE
. 7
EV
Example covered later in Appendix B
36;
26;
26;
27;
33;
39;
39;
28;
35; !!1374357529; @@76;
2280.00;
917.00; !!21.61; !!27.86
No tag - Value is within normal operating range
^^ - Value is at first warning level (logging only)
@@ - Value is at second warning level(near limit – Entry placed in logfile.txt)
!! - Value is over the set limits (may cause shutdown or loss of scanning) Operator Pop-up
appears every 5 minutes
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
; Starting log system
; TempMonColumnHeadings(1920); TXPS Temp In;TXPS
; TempMon(3372);
40;
24;
39;
29;
26;
25;
23;
29;
1712;
1945;
S
ROPRIETARY
27;
29;
26;
26;
28;
40;
29;
28;
32;
1988;
1955;
29.00;
M
ERVICE
ANUAL
29;
40;
25;
36;
34;
32;
26.00;
7 - 257