Electrical Problems - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Carestation
620/650/650c (A1)

Electrical problems

WARNING
Symptom
Mains indicator is not on.
One electrical outlet does
not have power.
A circuit breaker opens
frequently.
Tec
6 Plus vaporizer has
no power.
The real time clock on the
anesthesia display does not
maintain accurate time and
date.
Audible alarm sounds. The
system function stops. The
screen is blank.
Display does not function.
Task light does not function Task light is broken.
On/Standby switch does not
work
Touch screen does not
function.
7-18
If a circuit breaker opens frequently, do not use the
system. Contact an authorized service representative to
repair the system.
.
Problem
The electrical power cable is not
connected.
The power cable is damaged.
The electrical socket the power cable
connects to has no power.
An internal fuse is open.
The power management board does not
work. The cables between mains and the
power management board or the power
management board and the anesthesia
control board are disconnected.
The outlet circuit breaker is off.
The electrical outlet is damaged.
Equipment connected to the outlet(s) uses
more current than the circuit breaker
rating.
The equipment connected to the outlet
has a short.
Not plugged into outlet.
The outlet circuit breaker is off.
The battery in the anesthesia display
needs to be replaced.
Internal power converter failure.
Display is broken.
System does not have electrical or battery
power.
On/Standby switch is broken.
Touch screen is locked, dirty or broken.
Solution
Connect the power cable.
Replace the power cable.
Use a different electrical socket.
Contact an authorized service
representative to repair the system.
Contact an authorized service
representative to repair the system.
Turn the circuit breaker on.
Contact an authorized service
representative to repair the system.
Use a different power supply for some of
the equipment.
Contact an authorized service
representative to repair the system.
Connect power cable.
Turn the circuit breaker on.
Contact an authorized service
representative to repair the system.
Turn the system off. Contact an authorized
service representative to repair the system.
Contact an authorized service
representative to repair the system.
Contact an authorized service
representative to repair the system.
Confirm power cable is connected.
Contact an authorized service
representative to repair the system.
Make sure there is no liquid or other dirt on
the touch screen and that the screen is
unlocked. Contact an authorized service
representative to repair the system if the
problem continues.