Trends setup

2076152-001
Setting principal setting defaults
1.
From Super User mode, select System Config. and then
Parameter Settings.
2.
Select the item to change.
For Principal Volume, select MV or TV from the drop-down list.
For Principal O2, select Et or Fi from the drop-down list.
For Principal Agent, select Et or Fi from the drop-down list.
3.
Select Back to return to the System Config. menu.
Setting volume conditions and humidity
compensation
Note
When set to ATPD, the volume calculation conditions are based on
ambient temperature and pressure, dry humidity conditions. When
set to BTPS, the volume calculation conditions are based on body
temperature, ambient pressure, saturated humidity conditions.
1.
From Super User mode, select System Config. and then
Parameter Settings.
2.
Select the item to change.
To set the volume calculation conditions, set Vent and
Airway Gas Module TV Basis. Set to ATPD/ATPD, ATPD/
BTPS, or BTPS/BTPS.
To set the humidity compensation type, set CO2 Numbers
to Dry or Wet.
3.
Select Back to return to the System Config. menu.
Setting inspiratory flow
1.
From Super User mode, select System Config. and then
Parameter Settings.
2.
Select Insp Flow to set the graphical display for flow data.
Select the plus symbol to graphically display inspiratory flow as
a positive value.
Select the minus symbol to graphically display the expiratory
flow as a positive value.
3.
Select Back to return to the System Config. menu.
Use the Trends Setup menu to set the default trend graphical page
layout for the top, middle, and bottom trends for each page shown in
the user's Trends menu.