Chapter 5 Electrical Requirements; Power Requirements; System Electrical Characteristics; Core System - GE Optima IGS 320 Preinstallation Manual

Table of Contents

Advertisement

Chapter 5 Electrical Requirements

1 Power Requirements

1.1 System Electrical Characteristics

NOTICE
It is the customer's responsibility to ensure that electrical installation is compliant to local
regulations. To avoid risk of electric shock, this equipment must only be connected to a supply
mains with Protective Earth.
Power supply and ground cables shall be dedicated to the system. They must not be used to
supply other systems.
Power supply and ground cables shall be kept separated from room system cables.
Power supply and ground cables must be connected to the same distribution panel. They must
run near one to the other.
Power supply and ground cables provided by the customer shall be compliant with local
regulations (e.g. UL, NFPA 70, CSA, IEC, CCC).

1.1.1 Core system

Table 5-1:
Nominal voltage
380 V ± 10%
400 V ± 10%
415 V ± 10%
480V ± 10%
The Hospital circuit breaker should fit the current protection of the system:
150A/ 480V, 3 Phases for UL
80A/ 380V, 3 Phases for CE
Chapter 5 Electrical Requirements
Frequency
50 Hz or 60 Hz (± 3 Hz)
60 Hz only (± 3 Hz)
WARNING
THIS FLUORO UPS IS ONLY DESIGNED TO OPERATE IN A WYE-
CONFIGURED ELECTRICAL SYSTEM WITH A SOLIDLY GROUNDED
NEUTRAL.
FOR MORE DETAILS, REFER TO THE FLUORO UPS DOCUMENTATION.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Power consumption
Nominal
Peak
60 kVA
150 kVA
Direction 5537562-1-1EN, Revision 3
Type of power input
Without Fluoro
UPS option
3~
With Fluoro UPS
option
3N~
163

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima igs 330

Table of Contents