Ventilation modes factory default settings

Setting
Range
TV
20-1500 ml
RR
4-100 bpm
RR
2-60 bpm
Minimum RR
4-60 /min
I:E
2:1 - 1:8
Tpause
Off, 5-60% of
Tinsp
PEEP
Off, 4-30
cmH2O
Pmax
12-100
cmH2O
Pinsp
5-60 cmH2O
Psupport
Off, 2-40
cmH2O
Tinsp
0.2-5 seconds *
2076152-001
Tinsp
Rise Rate
1
2
T
I
1.
Paw
2.
PEEP
3.
Time
Figure 11-14 • CPAP + PSV waveform
The ventilation mode factory default settings for the ADULT case
default show in the following table. An * indicates the setting is not
used for the ventilation mode. See the "Super user mode" section for
all other case default values.
.
Mode
VCV
PCV
PCV-VG SIMV VCV SIMV PCV SIMV
500
*
500
12
12
12
*
*
*
*
*
*
1:2
1:2
1:2
Off
*
*
Off
Off
Off
40
40
40
*
5
*
*
*
*
*
*
11 Specifications and theory of operation
3
500
*
*
*
12
12
*
*
*
*
Off
*
Off
Off
40
40
*
5
2
2
1.7
1.7
PSVPro CPAP
PCV-VG
+PSV
500
*
*
*
*
*
12
12
*
*
*
12
*
*
*
*
*
*
Off
Off
Off
40
40
40
*
5
5
2
2
2
1.7
1.7
1.7
11-35