2076152-001
6.
Replace the O2 cell. Install the O-ring onto the adapter and
screw the adapter onto the new O2 cell.
2
1
1.
O2 cell
2.
Adapter
3.
O-ring
7.
Reconnect the cable connector to the O2 cell and reinstall into
holder.
8.
Rotate up and push the breathing system base to lock it into
place. A click sound will be heard when attached properly.
9.
Close the breathing system door.
10. Place the canister onto the canister lifter base.
11. Pull the canister lifter handle up to lock the canister into the
breathing system.