Ventilator operating specifications

Pneumatics

Fresh gas compensation

Pressure

Volume

2076152-001
11 Specifications and theory of operation
Based on the ventilator settings in use, the setting ranges may be
constrained so that the specifications stated are met.
Gas source
Gas composition
Peak gas flow
Flow valve range
Flow compensation range
Gas composition
Patient airway pressure
measurement range
High pressure alarm set range
Sustained pressure alarm range
Display range
Tidal volume display range
Tidal volume setting range
Minute volume display range
Anesthesia system
Medical Air or O2
120 l/min at 240 kPa (35 psi), 0.75 seconds
1 to 120 l/min at 240 kPa (35 psi)
0.1 to 15 l/min
O2, N2O, Air, anesthetic agents
-20 to 120 cmH2O, +/- 1 cmH2O
resolution
12 to 100 cmH2O, 1 cm increment
6 to 40 cmH2O, 1 cm increment
-20 to 120 cmH2O
less than 1 to 9999 ml, 1 ml resolution
20 to 1500 ml
less than 0.1 to 99.9 l/min, 0.1 liter
resolution
11-37